MIL-STD-1330D(SH) APPENDIX D
w/ Change 1
DATA SHEET 3. Establishing Dish Acceptance Criteria.
EVAPORATING DISH:
Name and ID (e.g., badge): _Date:
Date evaporating dish was last calibrated:
Batch number for standard:
Solvent used for standard: (Shall be same as solvent used for cleaning the instrument.)
Baseline applies: Yes No ; if "yes" lot or batch #
Gain Settings: Sensor:
Control Unit:_
Clean Dish Acceptance Criteria
Reading and Average |
SQM Readings |
|
Cleaning #1 |
Cleaning #2 |
|
Reading #1 |
||
Reading #2 |
||
Average for Reading #2 1/ (Clean Dish Acceptance Criteria) |
||
NOTE: 1/ For reading #2, the difference between the readings shall be less than 100 counts. |
Dish 5 ppm Acceptance Criteria
Reading/Dose and average |
SQM readings |
||
Trial 1 |
Trial 2 |
Trial 3 |
|
Cleaned Dish |
|||
Reading #1 |
|||
Reading #2 2/ |
|||
Dish 5 ppm Acceptance Criteria |
|||
Dose #3 |
|||
Average Dose #3 3/ (Dish 5 ppm Acceptance Criteria) |
|||
NOTES: 2/ For all trials, for reading #2, the difference between the clean dish acceptance criteria and each result must be less than 100 counts. 3/ For dose #3, calculate the difference between the highest and lowest readings. If it meets the acceptance criteria in Table D-III or D-IV, as applicable, the data is acceptable. If it does not meet the acceptance criteria, perform additional trials until results are acceptable. |