MIL-STD-1330D(SH) APPENDIx D
w/ Change 1
DATA SHEET 1. Surface Quality Monitor maintenance record for sensor electronics.
Date |
Name and ID |
Sensor Gain Setting used for establishing dish acceptance criteria from D.S. 3 (e.g., 4) Setting: (see D.8.3.1.2) |
Increase Sensor Gain Setting by 1 unit (e.g., 5) Setting: (see D.8.3.1.5) |
Decrease Sensor Gain Setting by 1 unit (e.g., 3) Setting: (see D.8.3.1.6) |
|||
SQM Reading |
Acceptance Criteria = SQM reading x 0.95 |
SQM Reading |
Acceptance Criteria = SQM reading x 0.95 |
SQM Reading |
Acceptance Criteria = SQM reading x 0.95 |
||
Instructions for filling out data sheet: 1. Enter date, name, ID. Enter the Sensor Gain Settings that will be used. 2. The first set of readings is entered in the first row. This is used to calculate acceptance criteria for subsequent readings. There are no acceptance criteria for these first readings. 3. For subsequent readings, compare the SQM reading to the acceptance criteria in the first row. If the SQM reading is less than the acceptance criteria, send the entire unit to the manufacturer for servicing. 4. If the first reading needs to be reestablished (e.g., if maintenance is performed by the manufacturer), start a new sheet. |
75