MIL-STD-1330D(SH) APPENDIX C
20 September 1996
C.8 Preparation and maintenance of apparatus.
C.8.1 When handling a NAVSEA approved halogenated solvent, the safety precautions specified by the material safety data sheet (MSDS) and 4.4.1 shall be followed.
C.8.2 Apparatus used in the procedure, including the sample bottles, homogenizer, homogenizing container, graduated cylinder, filtration reservoir, and demineralized water container, shall have been cleaned using a component cleaning process as specified in this standard, or equal.
C.8.3 After contact with NOC or TSP cleaning solution, all apparatus shall immediately be rinsed with demineralized water to remove residual cleaner, to prevent formation of inorganic deposits and to maintain cleanliness.
C.9 Procedure.
C.9.1 Cool the 500 to 600 mL NOC or TSP sample to room temperature or less. Acidify TSP samples to a pH of 2 or less.
C.9.2 Homogenize the 500 to 600 mL sample for 3 minutes at a speed of about
7,500 rpm (position 3 on an OMNI Model GLH homogenizer).
NOTE 1: Any NAVSEA approved halogenated solvent can be used for post NVR analysis. Only HCFC-225s and CFC-113 solvent can be used for post analysis by infrared spectrophotometry.
C.9.3 Transfer 100 mL or 500 mL of homogenized sample to the separatory funnel.
C.9.4 Transfer 160 mL of a NAVSEA approved halogenated solvent or 30 mL of HCFC-225s or CFC-113 to the separatory funnel. Cap and vigorously agitate for 30 seconds. Allow to stand until contents settle and bubbles disappear. Remove the cap carefully to release any pressure build-up. Drain solvent from separatory funnel and transfer to volumetric flask. Repeat two additional times.
C.9.5 Add fresh solvent to the volumetric flask and dilute solvent extract to 500 mL if serially extracted with 160 mL volumes or to 100 mL if serially extracted with 30 mL volumes. Shake to completely mix the solvent extract. Remove the cap carefully to release any pressure build-up.
C.9.6 Perform an NVR analysis of the 500 mL solvent extract as specified in
5.13.2.5, or perform a hydrocarbon oil and grease analysis of the 100 mL solvent
extract as specified in 5.13.2.6.
67/68